Neurotrac Pelvitone specifications:
1. 1. Dual channel: individually isolated circuits.
2. 2. Amplitude: 0-90 mA into 500 Ohm load; actual mA will tend to be less than indicated due to electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 86 mA, at 1500 Ohms load the maximum will be limited to 65 mA.
3. Type: Constant current,
maximum output voltage 180 Volts +10 / -30 Volts
4. Waveform: Symmetrical, rectangular bi-phasic with zero DC
current.
5. Selectable pulse width: 50 ìS 330 ìS [2% accuracy].
6. Pulse Rate selection: in the continuous mode 2 100 Hz
[2% accuracy].
7. Time duration of the treatment selectable: 1 minute to 90 minutes.
8. Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2
volts the battery symbol will flash on/off once every second.
9. Open Electrode Detect: If an open circuit is detected at the
output of channel A or B the output current will be reset at zero.
10. Ramp up time 0.1 - 9.9 seconds.
11. If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not turn
on.
12. Physical dimensions: 134 x 69 x 29.7 mm.
13. Weight: 0.18 KG with battery.
14. Environmental Conditions for use:
+10 to +30 degrees Centigrade. 0-90% Humidity.
15. Environmental conditions for storage & transport:
-10 to +50 degrees Centigrade. 0-90% Humidity.