Neurotrac MyoPlus 4 Specifications
1. Neurotrac MyoPlus 4 EMG Specifications:
" Dual Channel EMG
" EMG Range: 0.2 - 2000 µV RMS
" Sensitivity: 0.1 µV RMS
" Accuracy: 4% of µV reading +/- 0.3 µV at 200 Hz
" Selectable Bandpass filter - 3db Bandwidth,
b) Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% -Reading below 235 microvolts
10 Hz +/- 3 Hz to 370 Hz +/- 10% - Reading above 235 microvolts
b) Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
" Notch Filter: 50 Hz (Canada 60 Hz) - 33 dbs (0.1% accuracy)
" Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
" Battery: 6 V Ni-MH Rechargeable Model AR110-2300
Charger: Model ARBC-110; Input: 12V DC, 1.0A; Output: 6.0V DC, 0.7A
" Work/Rest periods: 2 - 99 seconds
" Number of Trials: 1 - 99
2. Neurotrac MyoPlus 4 STIM (Neuromuscular Stimulation) Specifications:
" 4 Channel Stimulator
" Amplitude: 0-90 mA into 500 Ohm load-actual mA will tend to be less than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 75 mA, at 1500 Ohms load the maximum will be limited to 50 mA
" Type: Constant current, maximum output voltage 70 Volts + 5 / - 10 Volts
" Waveform: Symmetrical, rectangular, bi-phasic with net zero DC current
" Pulse width selection: 50 - 450 µS (2% accuracy)
" Pulse rate selection: 2 - 100 Hz (2% accuracy)
" Work / Rest periods: 2 - 99 seconds
" Time 1 - 99 minutes
" Ramp up time: 0.1 - 9 seconds
" Preset and user programmable treatment Programs
" Automatic output shut off with detection of open electrode above 0.5 mA
3. Neurotrac MyoPlus Other Specifications:
" Low battery indication at 5 .2 volts +/- 0.2 volts
" Automatic shut off 4 minutes after pressing last key
" Environmental Conditions for use: + 10 to + 30 degress Centigrade with 0 - 90% Humidity.
" Dimensions: 150 mm long, 96 mm wide, 36 mm depth
" Weight: MyoPlus4 device 0.16 kg without battery / NiMH Alkaline battery: circa 0.2 kg
Neurotrac MyoPlus4 EMG Triggered STIM is a registered Medical Device with 2 years warranty.
Neurotrac Myo Plus4 summary overview:
" Dual channel EMG (Biofeedback) combined with 4 channels of NMS - STIM
" (Neuromuscular Stimulation) & ETS (EMG Triggered Stimulation).
" Passive and Active phases: EMG and NMS in one programme
" Manual and Automatic threshold control.
" EMG Biofeedback between stimulation (muscle activity and tiredness indication when using muscle stimulation).
" Stand alone device or used in conjunction with established multilingual software.
" Pre-set Clinical modes for: Continence, Sports, Rehab.
" Custom programmes enabling up to 5 phases (any of: STIM W/R
" -MOD-CONT-ALT, EMG, ETS) to be set up in each programme.
" Multilingual LCD screen and voice prompts for Biofeedback.
" For home users: Produces and prints (with the optional PC Software) the comprehensive patient progress reports on a day to day basis, providing average readings (2 channels) on work/rest, peak, onset and muscle release times, work and rest deviations and the average mA current.
" Used as a learning tool for both patient and clinician with emphasis on improving treatment techniques in the use of EMG and Neuromuscular stimulation.
" Multilingual LCD screen prompts.
" High capacity rechargeable battery and charger supplied (No mains power required).
" Lock function in Plus 4 enables to lock a selected programme and record all the EMG, ETS and STIM statistics during the day.
" After downloading the data from Myoplus 4 to NeuroTrac Software, Daily Progress Report of the Patient's Home Compliance is available to review or print out.
" Patient Database.
" Visualisation of ETS training.
" Select the Patient from the database, start a Clinical session or review the previous sessions.
" 2 channels of EMG.